NORTHERN ILLINOIS UNIVERSITY -
Department of Mechanical EngineeringLab:
Thermocouple CalibrationObjective:
To calibrate a thermocouple and find the corresponding curve-fit correlation.| Apparatus: Thermocouple; Temperature controllable bath; Digital Thermometer; Multimeter. Useful Links: Introduction to Thermocouples Temperature Meas. or [Eqs] Thermocouple Color Codes Technical Reference [K-Tble] |
Specifications: Thermocouple: Omega make, Type: J, K, or T thermocouples Constant temperature bath: Haake A81, 115V / 60Hz / 1500 VA Max temperature 180° C, precision 0.1° C Liquid : Water or Antifreeze circulated to maintain constant temperature. Digital Thermometer: Omega HH23 (K, J, T type); 5 digit LCD, 0.2° C repeatability (Similar product) Multimeter: Hewlett-Packard type 3478 or HP 34401A digital multimeter, 1 m V resolution; |

Theory: The thermocouples are based on the Seebeck effect, see
Sec.8.5-Thermoelectric Temperature measurement, Text p.343-362 or similar reference. The
Seebeck effect states when two dissimilar metal wire are connected with each other in a
loop to form two junctions, maintained at two different temperatures, a voltage potential
or electromotive force (E=emf) will be generated and the current will flow through the
loop circuit. The current will be proportional to the difference in temperature between
the junctions and the metals used. The higher the temperature difference, the higher is
the electromotive force (emf) and the current flow in the loop. The magnitude of the emf
is in the order of few millivolts. We need a precise and sensitive multimeter (MM) which
can read up to microvolts to do this experiment.
Procedure :
1. Check the level of water in the bath. Connect the apparatus to the Power supply.
2. Turn on the water bath by switching the main switch (1) as shown in the figure.
3. Keep the set-in switch (8) depressed and set the temperature to desired level (30 deg C to start) by turning the knob(6) and observing the display(7). Release the switch(8) after setting the temperature. In normal mode, the temperature shown on the digital display is the actual temperature of the bath (TB) against which the thermocouple sensor is to be calibrated.
4. Connect the ends of the thermocouple to the digital multimeter (MM) and set the multimeter to read in millivolts DC.
5. Dip one junction of the thermocouple in the thermo-bath liquid and wait for few minutes for it to reach the steady state (i.e. the reading on the MM steadies down except the last digit). Be careful to hold (tape) the sensor wire away from the circulator's propeller!
6. Note down the digital MM reading in millivolts (EMM) and repeat the steps 3 through 6 in steps of 5 degrees from 30 to 60 degrees Celsius.
7. As mentioned in the theory, the multimeter reading corresponds to the difference in temperature between the surroundings (room) and the bath. To calibrate the thermocouple, we have to take the room temperature into consideration to get the absolute value of temperature measured. Find the equivalent millivolt value for the room temperature from the corresponding Thermocouple table (ERM). Then, add that millivolt value, corresponding to the room temperature, to every multimeter reading (EMM). Tabulate the values.
Observations (Type-K thermocouple used):
Bath Temperature, TB [° C] |
Bath Temperature, TB [° F] |
MM reading, EMM [mV] |
Total emf, E = EMM+ ERM [mV] |
Temperature corresponding to E, TTC [° C] |
Temperature corresp. to E, TTC [° F] |
30 |
0.361 |
1.179 |
29.4 |
||
35 |
0.579 |
1.397 |
34.8 |
||
40 |
0.788 |
1.606 |
39.9 |
||
45 |
0.995 |
1.813 |
44.9 |
||
50 |
1.202 |
2.020 |
49.9 |
||
55 |
1.407 |
2.225 |
54.9 |
||
60 |
1.613 |
2.431 |
59.8 |
Plot the measured bath temperatures values (TB) on y-axis against the
corresponding thermocouple emf (millivolt) values (E) on x-axis. Find the slope, intercept
and the correlation coefficient of the curve-fitted line by any method. If the correlation
coefficient is not very close to one, curve fit with higher order polynomial. Calculate
the 95% uncertainty in temperature measurement of the calibrated thermocouple sensor if
the obtained curve-fit correlation is used. Worked out
Problem 5.30 should help you
with the assignment.
NOTE:
The multimeter voltage reading should be zero when you measure room temperate without the
reference junction (the thermocouple connected directly to the meter) Then the reference
temperature is the same as measured temperature, both equal to the room temperature,
therefore the temperature difference and generated voltage should be zeros. However, your
reading will never be exactly zero due to many reasons: