Experimental Methods Tips, Questions & Answers (Q & A)
Q: Are the Lab Handouts posted on the Web sufficient for
the lab assignments?
A:
Q:
Will a multimeter read zero Volts if it is not connected to anything?Q: If a
thermocouple is connected to a precise voltmeter (both being in a room), will it read
"zero" voltage corresponding to the zero temperature difference of its
junctions? Q: If a
thermocouple (being in a room) is connected to a precise voltmeter, will it read the table
voltage corresponding to the room temperature?
A: The
multimeter voltage reading should be zero when you measure room temperate without the
reference junction (the thermocouple connected directly to the meter) Then the reference
temperature is the same as measured temperature, both equal to the room temperature,
therefore the temperature difference and generated voltage should be zeros. However, your
reading will never be exactly zero due to many reasons:
A: Yes and No! Yes, if you have the reference junction in the ice-bath,
since the table values are given for the zero reference-junction temperature. However, if
only one-junction thermocouple (which is often the case) is connected to a voltmeter, the
voltmeter circuitry will make up the other (reference) junction. Since the both junctions
are at the same (room) temperature the reading should be zero volts, except for
measurement errors. The reference junction is actually separated by the meter's circuitry,
but as long as the meter terminals and circuitry are at the uniform temperature (any
level) and accounted for with the corresponding emf value for that temperature, it should
not matter, according to the rule of "intermediate metal interference."
Q:
How long does it take for a process to come to the steady state? How do we know it?Q: What is difference between an instrument resolution and accuracy and how repeating measurements and statistics relates to them?
Resolution is the output scale precision, to what smallest part we are able to "resolve" an instrument reading. If it is a digital scale, it is the last digit and instrument rounds the reading to the closest displayed digit for us. If it is an analog scale, than it is the smallest fraction we could visually resolve the indicator position against the scale divisions, regardless whether they are marked or we imagine them (then somewhat subjective).
Accuracy of an instrument is separate from its display resolution and depends on quality of instrument components and calibration. For example, a scale may have display resolution in grams and systematically be off for say 10 grams - that is accuracy.
In addition to resolution and accuracy, when we repeat the measurements we usually get variability (scatter) in repetitive readings due to inevitable changes in time of instrument properties, measurement procedure and/or measured variable. Statistics helps here by describing and averaging ("smoothing") those variabilities. However, if the variabilities are smaller than the instrument resolution we may never detect and be aware of it. For example, when we measure a battery voltage with large instrument resolution (say 0.1V) the measurement reading appears very steady, but if we use a very precise voltmeter (say 0.0001V resolution) then we may be measuring (detecting) very dynamic fluctuation, which may be due to many things, like instabilities in instrument properties, measured signal, and/or interference and noise from the surrounding (regardless that we are displaying 0.0001V the accuracy may be off for 0.01V for example) - and that is always the case. So, the statistics (of many repetitive measurements) does not help with systematic instrument errors, the latter may be corrected with proper (re)calibration. Hope this will not confuse and be of help.
Q: Will
A: Not